GE
DRAFT
LOGIQ P9/P7
D
IRECTION
5604324
, R
EVISION
11
DRAFT (J
ANUARY
24, 2019)
S
ERVICE
M
ANUAL
Chapter 7 - Diagnostics/Troubleshooting
7-13
7-4-3-5-3
Log message of System error
Log message format
“Error ; EAErrorHandler(****); Error from” + ErrorModu ErrorMessage
7-4-3-5-4
Types
7-4-3-5-4-1
Example1 Hardware Errors:
Please see the below tables for the details.
Exception Types:
Sub Type Codes:
GFI Hardware Exception
GFI Setup Exception
GFI Buffer Exception
GFI Processing Exception
Code
Error Name
Description
31
eSwHardwareException_ErrorHandler_GFIHardwareError
Hardware Error (MCB/ MDC/MPI..)
167
eSwProcessingException_TsLevelOk_Timeout
HV (High Voltage) is not available from MDC.
Hardware error
“Error ; EAErrorHandler(3276); Error from HouseKeeping : GFI Hardware Exception:
SubType 31: -GFE Ch0 empty-GFE Ch1 empty-CPS_ERROR(FAULT_HVLV). Severity :
FatalHW”
Error Module Name
Error descrition
Exception Type
Sub Type Code
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...