GE
RAFT
LOGIQ P9/P7
D
IRECTION
5604324
, R
EVISION
11
DRAFT (J
ANUARY
24, 2019)
S
ERVICE
M
ANUAL
6-6
Section 6-3 - LCD Monitor Adjustment
6-3-3
21.5 inch Wide Monitor Adjustment (R2.x.x and later)
All Monitor settings have been optimized at the factory, so normally, there should be no need for any
further adjustments.
1) Go to Utility second touch panel
2) Adjust Room profile, Color Profile, Gamma, Brightness/Contrast (Roomprofile-User defined) as
neccesary.
Figure 6-4 Monitor Adjustment
Display Test patterns
while the Utility is active
Select Color Profile
Select Gamma
Display scan screen
while the Utility is ac-
tive
Select Room Profile
Adjust Brightness and Constrast for Room profile
“User Defined”
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...