Chapter 4 - Functional Checks
4-30
4-3-7-6
M-Mode
1) Press B Mode Button.
2) Press M Mode Button. (Active M Mode button on "User Configurable Mode Rotary".)
3) Verify that system scan B/M with the activated probe.
Figure 4-29 M-Mode
4-3-7-7
B-Flow Mode
NOTE:
Make sure to select probe which supports B-flow
1) From the touch panel, select Utility -> Imaging -> General.
2) Change BF/BFC Button to “BF”
Figure 4-30 BF Setting
3) Press B-Flow Button. (Active M Mode button on "User Defined Hard Key".)