GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
3-14
Section 3-4 - Preparing for Installation
3-5-3
Transducer Connection
1.) Connect a transducer to one of the three rightmost transducer receptacle as follows:
A.) Ensure that the transducer twist lock lever to the horizontal position.
B.) Insert the transducer connector on the receptacle guide pin until it touches the receptacle
mating surface.
C.) Twist the transducer twist lock lever to horizontal position to lock it in place. Twist the lever
to the vertical position to disconnect the transducer.
NOTE:
It is not necessary to turn OFF power to connect or disconnect a transducer.
Содержание LOGIQ P6
Страница 2: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 2 ...
Страница 9: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SEVICE MANUAL vii JA ZH CN KO ...
Страница 22: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 10 ...
Страница 130: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL 4 50 Section 4 9 Board Function Checks ...
Страница 477: ......