Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-33
SYSCON PM Assy/SYSCON CM Assy
8-2-33-1
Tools
•
Common pilIips screwdrivers
8-2-33-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-33-3
Removal procedure
1.) Remove the Side Right Cover. Refer to the
8-2-16 "Side Right Cover" on page 8-52
.
2.) Remove the EMI Cover R. Refer to the
8-2-27 "EMI Cover R" on page 8-69
.
3.) Remove the SYSCONPM(SYSCONCM) assy in slot #3 in the nest box.
4.) Eject one ASIR ROM from old SYSCONML with regular screw driver.
Figure 8-116 PWAs in Nest Box