GE H
EALTHCARE
D
IRECTION
5394141, R
EVISION
5
LOGIQ™ P5 S
EVICE
M
ANUAL
Section 4-9 - Board Function Checks
4-49
4-9-6
P3RLY,P2RLY and P2RLY with Dummy Assy function check procedure
1.) Turn on the system.
2.) After completion of echoloader loading, connect probe on port 1. Refer to
3.) Log into “Service Login”. Refer to
5-10-2-2 "Service Login" on page 5-41
.
4.) Click “Diagnostics” on top menu.
5.) On left folder window, select 'LOGIQ P5 Diagnostics' => 'FRU test' => 'P3RLY'
6.) Select 'P3RLY Overall' and execute test
7.) All diagnostic test items must be passed.
8.) Shut down the system.
9.) Restart the system
10.)Activate probe. Refer to
4-3-7-2 "Activating the probe" on page 4-12
.
11.)Check the system basic functions. Refer to
4-3-4 "System B/M-Mode Checks" on page 4-8
and
4-3-5 "System CFM and PWD Checks" on page 4-9
.
-
For LA5 system, omit CFM test
12.)Disconnect probe. Refer to
4-3-7-4 "Disconnecting the probe" on page 4-12
.
13.)Connect probe on port 2 and repeat step 8) ~ 10)
14.)Connect probe on port 3 and repeat step 8) ~ 10)
-
For LA5 system, omit port 3 test
15.)All basic functions must show correct operation as described.
Figure 4-48 P3RLY diagnostic test