Chapter 9 Renewal Parts
9 - 43
* When replacing the R5 Upper Assembly, you may need to reload the C partition (only) in order to
have Windows set up the Touch Panel video position under the main screen.
Table 9-21 Upper Operator Panel Compatible sub FRU configurations for
5209000 family
If replacing
Op Panel – Upper
Sub-FRU(s)
:
Order
:
Hardware Compatibility
Software
Compatibility
Upper Bezel
5207000-15
Compatible with 5207000 and 5209000
families.
R5.x.x or later
Upper Operator
Panel Main
Control Board
5497858-11
Include cables for USB Cables to user
USB Ports.
Replaces 5497858-10
Compatible with 5209000 Family.
LCD and Frame
5497858-20
Compatible with 5209000 family.
Aux USB Board
5497858-30
This assembly part includes the USB
Ports and Cables.
Compatible with 5209000 family.
OP Cable Kit
5497858-40
Includes upper and Lower Cables.
Compatible with 5209000 family.
R5 Upper
Operator Panel
(complete OP Panel)
5209000-2
Replaces 5209000-1
Consoles 5205000-8 or later.
Only compatible with the R5 Upper Operator
Panel 5209000 family.