LOGIQ E9 S
ERVICE
M
ANUAL
7 - 54
Section 7-5 - Common Service Desktop
7-5-8-10
I/O Board Tests
7-5-3 "Diagnostics Window Overview" on page 7-16
for a description of the looping and list
features.
Symptoms:
•
Intermittent Problems
•
Popup messages warning of system voltage problems - I/O Board Pwr Supply Test
•
System comes up in simulator mode - run I/O Board GFI Voltage Test
•
Unable to run diagnostics
•
No doppler audio/system audio/general audio sounds - run I/O Board Audio Test
•
No display on the monitor
1.) Select the
I/O Board Tests
checkbox to run all the subtests, or select the individual subtests to run
only the selected tests.
2.) Enter the desired Loop Count (numeric value of 4 digits or less).
3.) Click Execute to run the test(s).