Chapter 7 Diagnostics/Troubleshooting
7 - 3
7-3-3
Collecting a Screen Capture with Logs
NOTE:
In R6 the screen capture is no longer taken.
If the system malfunctions, press the Alt+D keys simultaneously. This Alt+D function is available at all
times, and collects a screen capture of the image monitor, user-defined presets, and the following logs:
•
Keyboard Shadow Log (restricted in R6 or later)
•
Error Logs
•
Crash Log (restricted in R6 or later)
•
Vital Product Data
•
DICOM Logs
•
Windows Event Logs
•
Diagnostic Logs
•
Service Logs
For a detailed list of Service Logs captured,
see:
7-3-4 "Capturing Service Logs with ALT+D" on page 7-5
.
When Alt+D is pressed, a menu box opens. Enter the following information:
•
System ID serial number.
•
Software version.
•
System Date and time of occurrence.
•
Sequence of events leading to issue.
•
Is the issue repeatable?
•
Imaging mode, probe, preset/application.
•
Media brand, speed, capacity, type.
•
Select the Destination (storage media or
Service
directory for remote viewing through InSite ExC)
and click the Store button.
Figure 7-1 ALT+D Dialog Box
R5 and earlier
R6 and later