GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
8-10
Section 8-3 - Loading the System Software
8-3-2
Loading the System Software with USB Memory Stick
NOTE:
Before starting this procedure, remove all probes and peripherals and remove them from the
docking cart.
NOTE:
Please ensure AC adapter is connected during system upgrade!
1.) Properly turn off the scanner by momentarily pressing the
Power On/Off
Switch. Select “Full
Maintenance Reboot”/”Full Reboot” from the System Exit menu.
2.) If the system will not shutdown normally, hold down the
Power On/Off
Switch until the light
turns from green to amber.
3.) Insert the USB memory stick labeled “System & Application Software” to the system.
4.) Turn on the scanner. System will detect the USB memory stick automatically.
5.) Press any key to continue when below message display as shown in
.
Figure 8-12 Shutdown Dialog Box
Figure 8-13 System Software load message