Biopsy Procedures
301
LOGIQ
t
α
100 User Manual
2211157–100 Rev 0
Biopsy Needle Path Verification
To verify that the path of the needle is accurately indicated with-
in the guide zone on the system monitor, perform the following:
S
Properly install the bracket and biopsy guide.
S
Scan in a container filled with water.
S
Display the biopsy guide zone on the monitor.
S
Ensure that the needle echo falls within the guide
zone markers.
NOTE: The guide zone center line can be displayed by press-
ing the Measurement key.
The Procedure
S
Place coupling gel on the scanning surface of the probe/
sheath/biopsy guide assembly.
S
Invoke the biopsy guidelines by pressing Ctrl B.
NOTE: The selected biopsy zone angle designation is
displayed on the left hand side of the image display.
S
Scan to locate the target. Center the target in the electron-
ic guideline path.
S
Place the needle in the guide between the needle barrel
and needle clip.
S
Press the
MEASUREMENT key, to invoke the injection
depth. A dotted line appears in the middle of the Zone and
an open ended “ ” cursor appears on this line. The posi-
tion of the cursor gives the injection depth which is dis-
played on the left side of the image display. (Refer to page
122 for details of Biopsy Depth Measurements.
S
Use the
TRACKBALL to move the cursor along the center
line. The injection depth values are computed and dis-
played on the left side of the image display.
Post Biopsy
When the biopsy is complete, remove the needle barrel,
needle clip and probe sheath. Properly dispose of these
items in accordance with current facility guidelines.
The biopsy bracket can be sterilized and reused.
.
.
Содержание LOGIQ A100
Страница 18: ...Introduction LOGIQt α100 User Manual 2211157 100 Rev 0 2 This page left blank intentionally ...
Страница 30: ...Getting Started LOGIQt α100 User Manual 2211157 100 Rev 0 14 This page left blank intentionally ...
Страница 42: ...Preparing the System for Use 26 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 54: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 38 This page left blank intentionally ...
Страница 62: ...Patient Safety 46 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 74: ...Device Labels 58 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 76: ...Safety LOGIQt α100 User Manual 2211157 100 Rev 0 60 This page left blank intentionally ...
Страница 80: ...Patient Registration 64 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 88: ...Image Display 72 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 100: ...VCR Operations 84 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 114: ...General Measurements LOGIQt α100 User Manual 2211157 100 Rev 0 98 This page left blank intentionally ...
Страница 140: ...Diagnostic Category LOGIQt α100 User Manual 2211157 100 Rev 0 124 This page left blank intentionally ...
Страница 146: ...OB LOGIQt α100 User Manual 2211157 100 Rev 0 130 This page left blank intentionally ...
Страница 148: ...Exam Preparation 132 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 188: ...OB Measurements 172 LOGIQt α100 User Manual 2211157 100 Rev 0 This page intentionally left blank ...
Страница 224: ...OB Report Page 208 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 236: ...Cardiology 220 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 248: ...Urology LOGIQt α100 User Manual 2211157 100 Rev 0 232 This page left blank intentionally ...
Страница 251: ...LOGIQt α100 User Manual 2211157 100 Rev 0 235 Control Keys Configuration Using Control Keys ...
Страница 252: ...Control Keys LOGIQt α100 User Manual 2211157 100 Rev 0 236 This page left blank intentionally ...
Страница 272: ...Configuration Using Control Keys 256 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 274: ...Probes Biopsy LOGIQt α100 User Manual 2211157 100 Rev 0 258 This page left blank intentionally ...
Страница 298: ...Probes 282 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 326: ...Troubleshooting and Maintenance LOGIQt α100 User Manual 2211157 100 Rev 0 310 This page left blank intentionally ...
Страница 344: ...Maintenance 328 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 346: ...OB Tables LOGIQt α100 User Manual 2211157 100 Rev 0 330 This page left blank intentionally ...
Страница 400: ...OB Tables 384 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...
Страница 408: ...Index Index 8 LOGIQt α100 User Manual 2211157 100 Rev 0 This page left blank intentionally ...