Section 8-46 - Upper and Lower Column Cover Replacement Procedure
Section 8-46
Upper and Lower Column Cover Replacement Procedure
8-46-1
Manpower
2 persons, 3 hours
8-46-2
Tools
•
Hexagon key, 10 mm, with torque/momentum controller
•
Hexagon keys, 4,5 and 6 mm
•
Phillips screwdriver, size 1 and 3
8-46-3
Preparations
1.) Shut down the system as described in Power Off / Shutdown on page 4-3.
2.) Move Top Console to down position.
3.) Unscrew 4 screws on the Upper Column Cover.
4.) Remove monitor as described in
Monitor Assembly Replacement Procedure on page 8-92
5.) Remove Frogleg as described in
Frogleg Replacement Procedure on page 8-159
8-46-4
Upper and Lower Column Cover Removal Procedure
1.) Unscrew 2 Phillips screws on each side of Upper Column Cover.
2.) Lift Frogleg off its axle.
3.) Remove column covers.
8-46-5
Upper and Lower Column Cover Installation Procedure
1.) Thread on replacement cover(s) in correct order.
2.) Install Frogleg described in
Frogleg Installation Procedure on page 8-161
.
3.) Fasten 4 screws on Upper Column Cover.
4.) Install Monitor as described in
Monitor Replacement on page 8-94
Figure 8-172 Upper and Lower Column Cover
Screws
Upper Column Cover
Lower Column Cover