Chapter 8 Replacement Procedures
8-151
Section 8-37
Front Caster Replacement Procedure
8-37-1
Manpower
1 person, 30 minutes
8-37-2
Tools
Torx screwdriver T-20
Open End wrenches, 10 and 134 mm
5 mm hexagon key
Bevel Edged Board and Wooden Wedge (included in package with replacement Casters)
8-37-3
Preparations
1.) Power Down/Shutdown the system as described in
2.) Remove the Front Bumper as described on
3.) Turn Front Casters as shown in
.
4.) Activate Direction Lock.
5.) Put the Bevel Edged Board on the floor.
6.) Pull the system backwards up the board incline.
Figure 8-158 Bevel Edged Board and Wooden Wedge.
WARNING
WARNING
Prior to elevating scanner, verify that the floating keyboard is locked in its lowest
position.
WARNING
WARNING
Use extreme caution as long as VIVID 7 is un-stable, not resting on all four Casters.
Wooden Wedge
Bevel Edged Board