GE M
EDICAL
S
YSTEMS
PROPRIETARY TO GE
D
IRECTION
2294854-100, R
EVISION
3
LOGIQ™ 9 P
ROPRIETARY
M
ANUAL
8-120
Section 8-30 - Trackball Assembly Replacement Procedure
Section 8-30
Trackball Assembly Replacement Procedure
8-30-1
Manpower
One person, 1 hour 30 minutes
8-30-2
Tools
•
Phillips screwdriver size 1
•
Socket 2,5 mm
8-30-3
Preparations
1.) Shut down the system as described in
.
2.) Remove the Operator Panel as described in
Section Section 8-27 on page 8-111
3.) Put the panel upside down on a soft cushion.
Figure 8-130 Trackball
Trackball assembly
Screws
Trackball unit
Connectors
Trackball housing
Screws
Operator Panel with Trackball assembly