Chapter 8 Replacement Procedures
8-19
Section 8-4
Back End Processor Power Supply Replacement
8-4-1
Manpower
One person, 1 hour + travel.
8-4-2
Tools
Long Phillips driver.
8-4-3
Preparations
(click here to view the cover removal video)
1.) Power Down/Shutdown the system as described in
2.) Remove the left- and/or right-side covers as shown in
3.) Remove the upper rear cover as shown in
.
4.) Remove the lower rear cover as shown in
8-4-4
Removing the BEP Power Supply
1.) Disconnect the power cord from the back of the Back End Processor (BEP).
2.) Remove the cover from the Back End Processor (BEP) by loosening the thumb screw on the back.
Figure 8-10 Power Cable and Cover Screw
Step 1
Step 2