Chapter 7 Diagnostics/Troubleshooting
7-15
Section 7-7 Acquisition Diagnostics
Individual selections run several diagnostics in a
Bottom-up Confidence-Building Order
and return
status information identifying any problems. The Acquisition Diagnostic menu is structured as follows:
7-7-1
Automatic Troubleshooting
- Runs all diagnostics and analyzes the results to isolate to the faulty FRU.
-
This is the
BEST
way to diagnose a problem with the system.
-
Takes less than
6
minutes for all tests to run.
-
Disconnect all probes from the scanner before executing the Automatic Troubleshooting
sequence.
The choices for Automatic Troubleshooting, shown in
, allows for running all
acquisition diagnostics under the selection Troubleshoot All Acquisition Hardware.
Figure 7-15 Customer Acquisition Diagnostics