Optimizing the Image
5
-
60
LOGIQ 7 Online Help
Direction 2392536-100 Rev. 1
LOGIQView
Description
LOGIQView provides the ability to construct and view a static 2D
image which is wider than the field of view of a given transducer.
This feature allows viewing and measurements of anatomy that
is larger than what would fit in a single image. Examples include
scanning of vascular structures and connective tissues in the
arms and legs.
LOGIQView constructs the extended image from individual
image frames as the operator slides the transducer along the
surface of the skin in the direction of the scan plane. The quality
of the resulting image is somewhat user-dependent and requires
some additional skill and practice to develop proper technique
and become fully proficient.
LOGIQView is not available for the following:
• Multi
Image
• Timeline
Modes
•
B Flow Mode
•
Color Flow Mode
• PDI
Mode
Benefits
The user can look at a larger region of interest within one field of
view that is wider than any given probe would normally provide.
Clinical Use
LOGIQView is intended for scanning areas too large to fit on one
image.
Using LOGIQView
To perform an exam using LOGIQView,
1. Perform a detailed examination of the anatomy/pathology.
Optimize parameters for tissue texture and visible window
PRIOR TO activating LOGIQView.
2. Press the LOGIQView key on the Control Panel.
3. To start acquiring the image, press ‘L’ (the left split screen
key).
HINTS
When you scan, scan slowly and in a uniform motion
lengthwise, end-to-end (with or against the probe orientation
marker). LOGIQView acquires images via leading edge vectors
(and does not acquire slices, as in CINE). The image is being
stored as you perform the scan and you can watch the
LOGIQView as it is being acquired.
Содержание LOGIQ 7
Страница 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Страница 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 400: ...General Measurements and Calculations 7 94 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 699: ...LOGIQ 7 Basic User Manual 14 1 Direction 2392206 100 Rev 1 Chapter 14 ReportWriter Describes how to generate reports ...
Страница 741: ...LOGIQ 7 Basic User Manual 15 1 Direction 2392206 100 Rev 1 Chapter 15 Recording Images Describes how to record images ...
Страница 779: ...Presets LOGIQ 7 Basic User Manual 16 3 Direction 2392206 100 Rev 1 Overview continued Figure 16 1 Utility Touch Panel ...
Страница 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Страница 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Страница 983: ...GE Medical Systems ...
Страница 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...