Additional Cardiac Calculations
LOGIQ 500 Basic Users Manual
2233658–100 Rev. 0
10–27
Mean PG
To measure Mean PG (Mean Pressure Gradient):
MVA
Mean PG
LV
Report
PHT
ET
Max PG
TRACE
AUTO
Figure 10–31. Cardiac Calculation Sub-Menu
Select MEAN PG from the Calculation Sub-Menu. Perform the
measurement described on
7–15. Set the first cursor on the
waveform trace starting point.
If Trace Auto is selected, the waveform is automatically traced
after the second cursor is fixed.
If Trace Auto is not selected, manually trace the desired portion
of the waveform.
Press Set to complete the the Mean PG measurement.
Automatic Calculation of Mean PG
If velocity was previously measured and calculated in Doppler
Mode, Mean PG is calculated from the velocity value when
Mean PG is first selected in the Calculation Sub-Menu.
The waveform is automatically traced. Velocity and Mean PG
are also displayed on the screen.
Содержание Logiq 500
Страница 102: ...Exam Applic Preset Selection LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 4 8 This page left blank intentionally ...
Страница 179: ...Mixed Mode Display Formats LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 5 78 This page left blank intentionally ...
Страница 206: ...Annotating an Image LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 6 28 This page left blank intentionally ...
Страница 238: ...General Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 8 12 This page left blank intentionally ...
Страница 317: ...Advanced Cardiac Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 10 38 This page left blank intentionally ...
Страница 329: ...Urology Basic Calculations LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 12 4 This page left blank intentionally ...
Страница 364: ...Advanced Recording Option LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 13 36 This page left blank intentionally ...
Страница 555: ...Assistance LOGIQ 500 Basic Users Manual 2233658 100 Rev 0 16 44 This page left blank intentionally ...