Chapter 8 Replacement Procedures
8 - 177
Section 8-11
Installing Base Software Load Image
8-11-1
Installing Base Software Load Image CD
NOTE:
While we do not believe it is necessary, It would not hurt to disconnect the system from the
network and remove all transducers.
1.) Power ON the scanner.
2.) Tap gently the
DEL
key until the following screen appears.Enter
jupiter
as a password if the system
request a password, then press the
Enter
key.
3.) The Setup Utility screen appears. Using arrow keys, Select
Advanced
Tab and select
USB
Configuration
and then press the
Enter
key. Verify that
High-Speed USB
is
Enabled
. If it is
disabled, using arrow keys, select
High-Speed USB
and press the
Enter
key and change it
4.) To exit USB Configuration, press ESC key
Figure 8-202 Enter Current Password
Figure 8-203 USB Configuration