GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8 - 171
8-10-5-4
Operator Login Password Back-up
The Operator Login Password will be deleted. So, write down the password if necessary.
8-10-5-5
Image Management Guide
NOTE:
Before upgrading to R4.x.x software, perform a recommended image backup procedure.
Save As
to View on any PC - Use this to save images (Dicom or Raw Dicom) in a computer-friendly
format (.avi or .jpeg) so you can view it on any PC.
Export/Import
Data/Images Between Systems - Use this to copy both patient data and images for
specified patient(s) from one system to another.
Save As (Saving Images to CD-ROM to View on Any PC)
To save images to the CD-ROM:
1.) Insert the CD.
2.) Go to Utility --> Connectivity --> Tools(Removable Media instead of Tools in case of BTO5 SW)
Select the Media as CD Rewritable.
3.) Press Format to format the CD-ROM.
4.) Select the image(s) to be saved to CD-ROM, press Menu, and select Save As.
Figure 8-190 Operator Login
Parameters
Descriptions
Operator Login Password
Содержание LOGIQ 5 PRO