GE MEDICAL SYSTEMS
LOGIQ
a
200 PROPRIETARY MANUAL
2138854
DIAGNOSTICS
2–12
REV 1
2–3–2 System Diagnosis (Continued)
TABLE 2–1
SYSTEM DIAGNOSIS TEST LIST
TEST NAME
TEST DETAIL
TEST TYPE
FAILURE CODE
SYSTEM TEST
AUTO
–
BEAM FORMER TEST
MANUAL
–
HV TEST
AUTO
211
ADC TEST
MANUAL
–
DSC FRONT/END TEST
AUTO
311, 312
DSC Interpolation test
AUTO
321
B MODE FUNCTION TEST
DSC coordination test
AUTO
322
DSC ALM test
AUTO
323
B MODE X–Y MEMORY TEST
AUTO
324
B MODE CINE TEST
AUTO
330~340
M MODE FUNCTION TEST
M–Mode Control Circuit
AUTO
341~344
Battery for back up
AUTO
411
SCSI controller
AUTO
412
MST TEST
RAM for Battery backup
AUTO
413
MST TEST
Real Time Clock
AUTO
414
Counter (8253)
AUTO
415
RS232C
AUTO
417
Graphic Controller
AUTO
419, 420
SYSTEM DRAM TEST
AUTO
416
Application ROM Check Sum
AUTO
418
OS ROM
AUTO
421
KEYBOARD TEST
MANUAL
–
PIOP TEST
AUTO
600
MONITOR TEST
MANUAL
–
POWER SUPPLY TEST
AUTO
801-6
Note
There are two kinds of failure codes in B–Mode Cine test. In case of single memory fault, the
failure code 330~332 is displayed. The other failure case, the failure code 336~337 is appeared
because one frame image data is saved in both memories.
Содержание LOGIQ 200
Страница 4: ......
Страница 8: ......
Страница 10: ...05 23 00 MAC Page 2 of 2 ...
Страница 28: ...05 23 00 MAC Page 2 of 2 ...
Страница 87: ...LOGIQ α200 ...
Страница 88: ...LOGIQ 200 PRO ...
Страница 144: ......
Страница 168: ...GE MEDICAL SYSTEMS LOGIQ 200 PRO Series PROPRIETARY MANUAL 2242594 DIAGNOSTICS 2 31 B MODE CINE TEST ILLUSTRATION 2 15 ...
Страница 190: ......
Страница 196: ......