4 - 10
Section 4-5 - Software Configuration Checks
Section 4-5
Software Configuration Checks
Section 4-6
Peripheral Checks
Check that peripherals work as described below:.
Table 4-10 Software Configuration Checks
Step
Check
Expected Result(s)
If Not Remedy
1.
Check Date and Time setting
Date and Time are correct
Adjust the Date and Time setting
2.
Check that Location (Hospital Name) is
correct
Location Name is correct
Re-enter the correct Location Name
Table 4-11 Peripheral checks
Step
Check
Expected Result(s)
If Not Remedy
1.
Press Freeze Key
to stop image acquisition.
2.
Press
RECORD
Key on the Control
panel
The image displayed on the screen is
printed on B&W Video Graphic Printer
3.
Press
SHIFT + RECORD
Key on the
Control Panel.
To image or report page displayed on
the screen is printed on the Parallel
Printer connected to the Parallel Port
of the System
4.
Press
SHIFT + STORE
Key on the
Control Panel.
To start the Image Transfer Operation
5.
Press
EXTERNAL VIDEO
Key on the
Control Panel
Enables an External Video Input
Playback on the LOGIQ™ 180
monitor. Press once again to return to
the scan mode