5.2
Electrotransfer with LARGE SEMIPHOR TRANSPHOR
UNIT and SEMIPHOR TRANSPHOR UNIT
Introduction
Electrotransfer with LARGE SEMIPHOR TRANSPHOR UNIT and SEMIPHOR TRANSPHOR
UNIT are performed at low current and low voltage, in approximately one hour.
This section describes how to run the electrotransfer using LARGE SEMIPHOR TRANSPHOR
UNIT and SEMIPHOR TRANSPHOR UNIT.
Recommended current settings
Use the graph to quickly find the maximum current setting for your gel size , or calculate
the gel area (cm
2
) and multiply it by 0.8 mA.
Gel size
(cm
2
)
Maximum
current (mA)
14 × 24
14 × 16
8 × 10
8 × 7
500
400
300
200
100
0
0
100
200
300
400
45
65
180
270
SEMIPHOR TRANSPHOR UNIT series Operating Instructions 29281777 AA
45
5 Operation
5.2 Electrotransfer with LARGE SEMIPHOR TRANSPHOR UNIT and SEMIPHOR TRANSPHOR UNIT