Maintenance, Cleaning and Troubleshooting (KISS™/KISS™ Multilead)
Troubleshooting
Possible error
:
Electrodes do not become attached by suction, insufficient
suction, or electrodes fall off
Causes
:
No or insufficient negative pressure due to:
• Kink in negative pressure tubing
• Negative pressure tubing connection to pump interrupted
• Filter disk (5) soiled
• Power supply interrupted
• Leak in negative pressure system
• Broken or failed regulator O-Ring
Remedies
:
Remove kink in negative pressure tubing, establish connection
to pump, replace filter disk, replace regulator O-Rings, connect
power supply.
Instrumentation test as in
“Equipment Check for Leaks in the
Complete System, Pump, Distributor, Electrode Leads and Suction
Cups” on page 20
to identify leak.
Possible error:
ECG leads yield poor recording
Cause:
Electrode leads, trunk cable shielding, or plugs interrupted
Remedy:
Measure resistance as in
“Testing Electrical Function (KISS™)”
on page 24
.
22745010H
KISS™ / KISS™ Multilead
39
Содержание KISS
Страница 14: ...Functional Description 14 KISS KISS Multilead 22745010H ...
Страница 15: ...2 Operating Controls and System Components KISS 22745010H KISS KISS Multilead 15 ...
Страница 18: ...Operating Controls and System Components 18 KISS KISS Multilead 22745010H ...
Страница 36: ...Testing instructions for KISS and KISS Multilead 36 KISS KISS Multilead 22745010H ...
Страница 40: ...Maintenance Cleaning and Troubleshooting KISS KISS Multilead 40 KISS KISS Multilead 22745010H ...
Страница 44: ...Spare Parts List KISS KISS Multilead 44 KISS KISS Multilead 22745010H ...
Страница 52: ...Schematics 52 KISS KISS Multilead 22745010H ...
Страница 53: ......