Page 15
Intellix
™
BMT 330 User Manual
BMT330UM
v1.0
20-Jun-16
After completion of the software setup of the Intellix™
BMT 330, remove the PC connection by following the
above steps in reverse, eg:
Remove the PC connection,
Reinstate the Alarm PCB (if applicable),
Unground the coaxial inputs (if applicable),
Close the Intellix™ BMT 330
cabinet,
Re-energise the AC supplies to the Alarm Relay
PCB.
Opening the Setup software
Launch the BMT Setup Software. At the
Communication
page ensure the BMT
Setup Software can establish communication with the Intellix™ BMT 330: setup the
correct connection and settings and confirm them by pressing the “
Ping
” button. A
“Success” message must be returned. Otherwise go to Communication, page 20.
Select the
Access
page; and select the required access level from the drop down
as appropriate (
Customer
is the default selection):
Customer
requires no password; you can navigate around the side tabs
immediately, though you are offered a restricted selection.
Service
requires you to sign in using the Service Password Of The Month
(Service POTM). Ensure POTM is active. Refer to the Intellix™ BMT 330
Service Manual for the service options available.
Factory
requires you to sign in using the Factory POTM. For internal use
only.