8 Room Speaker
The remote speaker provides the operator in the exam room, the indication of record exposure
sequence (same level information as provided by the control room control panel).
The speaker must be installed in the exam room out of the patient vicinity to give easy access to
the FE for maintenance operations and to medical staff to adjust the tone signal level and to be
able to see the power LED indicator.
Illustration 5-24:
WARNING
THE REMOTE SPEAKER KIT SHALL NOT BE INSTALLED IN THE PATIENT
VICINITY.
Patient Vicinity Definition : see
Patient Environment Equipment in
Innova
TM
IGS 620, Innova
TM
IGS 630 Preinstallation Manual
Direction 5694389-1-1EN, Revision 2