5 X-Ray ON Lamp Distribution
The fitter shall install an X-Ray On lamp in the Exam room, and in the control room if necessary,
outside the patient area, so that they are connected to the equipment through connection
terminals.
NOTE:
Innova systems with 21 & 31 cm detector provides 24 VAC for driving X-Ray ON line.
Hospital or contractor to provide 1.5 mm
2
(AWG 14) wires to connect to the system.
X-Ray ON lamp is active during X-Ray exposures.
WARNING
THE X-RAY ON LAMP MUST BE INSTALLED IN THE EXAM ROOM TO
CONFORM TO INTERVENTIONAL STANDARD IEC/EN 60601-2-43.
SIGNAL INDICATING THE X-RAY ON SHALL BE PERCEPTIBLE BY THE
OPERATOR IN ALL THE LOCATIONS DEFINED FOR THE PERSONNEL WHO
MAY RECEIVE SCATTERED RADIATION.
In the control room, an additional X-Ray on light must be installed if the console (VCIM)
indicator cannot be perceived by all the persons in the control room.
Illustration 5-21: X-Ray ON cabling between C2 cabinet and PDB
Innova
TM
IGS 620, Innova
TM
IGS 630 Preinstallation Manual
Direction 5694389-1-1EN, Revision 2
Chapter 5 Electrical Requirements
195