3.4 Wall Requirements
The C1 Cabinets, the C2 Cabinet, the PDB Cabinet and LD cabinet (optional) must be securely
fastened to the wall to prevent them from tipping.
The optional LD secondary monitor outside the patient room is mounted on wall. The swingout
arm that holds the LD additional monitor shall be mount according to the manufacturer mounting
manual, see
Articulating Arm Wall Mount Installation Manual in OEM manuals list.
An hooking point shall be provided in order to uplift the LD additional monitor at the swingout
arm level during its installation:
•
Hooking point characteristic: It must withstand up to 440 lbs (200 Kgs)
•
Hooking point position:
Illustration 2-66: Hooking point position
•
Recommended hooking point dimensions:
Illustration 2-67: hooking point dimensions
The position of the Swing out arm Center of Gravity (with weights) is given below.
Innova
TM
IGS 620, Innova
TM
IGS 630 Preinstallation Manual
Direction 5694389-1-1EN, Revision 2
142