Illustration 5-21: Fluoro UPS Cabinets (Optional) - UPS UL
The left, right or back side of the UPS cabinet can be positioned against the wall.
The front side of the UPS cabinet must be accessible for maintenance operation.
In front of the cabinet, the clear width of the service area to insure electrical safety shall be at least
0.9 m. In cases where 2 cabinets are installed face to face (both sides of the access way), the
clear width shall be at least 1.2 m.
Recommended minimum clearance between ceiling and top of the UPS should be 400 mm (16”)
for proper cooling air exhaust.
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5177951-4-1EN, Revision 1
Chapter 5 Physical Characteristics
119