![GE Innova 2100-IQ Скачать руководство пользователя страница 192](http://html1.mh-extra.com/html/ge/innova-2100-iq/innova-2100-iq_preinstallation-manual_4332925192.webp)
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5314528-2-1EN, Revision 1
16.3 PHYSICAL RUN – SYSTEM CORE DETAIL
Table 6-4:
MIS
number
Cable
Assembly
UL Style
Voltage rating
(V)
Max Voltage
carried (V)
Cable
diameter
(mm)
Connector type
Bigger Plug
size (mm)
Group n
1 (From C1 / C2 cabinets to LC positioner / Table)
27663A
2285897
2789
600
13.8
DB 11 pin
34.4
27664A
2380494
2463
600
8.3
Metrimate 6 pin
29
27665A
2331836
2463
300
9.2
Metrimate 6 pin
29
27666A
2352342
WATER HOSE
27667A
2252342
WATER HOSE
27468A
2258191
1019
600
9.1
Pre-stripping, ring terminal
12
27473A
2258191
1019
600
9.1
Pre-stripping, ring terminal
12
27670A
2231647
2789
300
HES 15 pin
27510A
2261135
WATER HOSE
27511A
2261135
WATER HOSE
27512A
2337267
FIBER OPTIC
27513A
2250247
2464
300
7
CEI 320 - HBL4570C plug
38.6
27514A
5139188
2789
30
7.3
DB 26 pin
41.2
27675A
2361531
Type CL2
150
8.2
Pre-stripping
8.2
27567A
2337165
2464
300
9.9
DB 15 pin
41.9
27660A
2258294
600
6.5
Pre-stripping
27661A
2258294
600
6.5
Pre-stripping
27662A
2350071
30
7
AMP 8 pin RJ45
27669A
2266532
2464
300
16.2
Circular connector
47.8
27671A
2337252
2789
30
10.9
Amplimite 100 pin
84.6
27673A
2337252
2789
30
10.9
Amplimite 100 pin
84.6
27672A
2337252
2789
30
10.9
Amplimite 100 pin
84.6
27674A
2266584
2464
300
10.6
DB 25 pin
56.6
Following cables are part of component
27586A
2260420
27722A
2363958
300
15
AMP 25 pin
55
27589A
2258207
Locaflex spec.
1311418A
11.7
HV plug
72
27590A
2115483
Locaflex spec.
1311418A
11.7
HV plug
72
Group no.2 (From C1 / C2 cabinets to Control Booth)
27676A
2350102
300
8.3
SCSI.50
60
27472A
2259592
4299
300
250
8 plug power strip CEI320 -
(HBL4570C plug - ring terminal)
39
27492A
2264834
1354
27493A
2337270
Category 5
150
5.5
RJ45
14.1
192
16 Physical Runs
Содержание Innova 2100-IQ