2.4 Injector Mounting Requirements
Table accessory rail load consideration:
The maximum load per table accessory rail is 40 kg (88 lbs) at 150 mm (0.49 ft) (60
N.m or 44.25 ft/lbs). Therefore:
•
Only light extra load not exceeding 5 kg (11 lbs) at 100 mm (0.33 ft) (i.e IV
pole with its accessories, pressure head...) is authorized on the same
table accessory rail as the injector.
•
Never install injector and radiation protection on the same table accessory
rail.
•
Typical installation on the front table accessory rail is Smart handle or
Smart box, Table Side System Control (TSSC), InnovaCentral/Touchscreen,
Table panning device and cables support.
•
If needed an optional rail can be installed at table foot end of the Omega
V table for other options.