2.3.3.5 2285053 – Through bolts with insert
Table 5-12:
Item
Name
Part #
Description
Quan.
Notes
Bolt, Hex
2296892
Through Bolt; M20 - 500-400
12 bolts for LC Positioner and 4 bolts
for Table
16 pc
Through Bolt mounting
method hardware
Washer, Flat
99142204
Special Flat Washer for Through
Bolts; one for each bolt
16 pc
same as above
Plate
2290941
Special Steel Spacer Plate; 4 in. x
4 in. (102 mm x 102 mm); one for
each bolt
16 pc
same as above
Nut, Hex
99141607
Hex Nut M20 STL galvanized two for
each bolt
32 pc
same as above
Grout assy
2285055
- 10 kg Powdered Mortar Ardex K-15
- RTV Silicon Rubber Adhesive
- 18 mm making tape adhesive
- Open cell foam
1 kit
Used in constructing LC
Positioner grout dam
Dowel
2290937
Wood Dowel; 24 mm diameter
12 pc
Water
Electric
Separator
2268647
Manufactured part introduced in
Positioner key hole to fit 1 electrical
conduit of 6”.
1 pc
Electrical and water conduits
separated inch size.
144