![GE Hydran 201i Скачать руководство пользователя страница 77](http://html.mh-extra.com/html/ge/hydran-201i/hydran-201i_instruction-manual_128390077.webp)
User Interface Operating Modes
Part 17997
Rev. 6, March 2011
4-5
Table 4-1 - Messages Displayed in the Main Display Mode
4.2.3
Menu Navigation Mode
The Menu Navigation mode is used to access the Main Menu or the Extended Menu, and
thus to all H201T
i
’s parameters and commands without requiring the use of a host
computer.
• The Main Menu gives access to the
most frequently used
parameters and commands.
• The Extended Menu gives access to
all
parameters and commands.
Note: The H201Ti automatically returns to the Main Display mode if the keypad is not used
for five minutes.
Each submenu of the Main Menu and Extended Menu is described in detail in Section 4.5
on page 4-10.
Display
Description
Alarm message
(
Alarm set point
)
Displayed only if an alarm occurred, if it was acknowledged and if the
alarm condition is still present. See Section 4.2.1 on page 4-3.
GasLevel
xxx PPM
• Gas level measured by the Hydran 201 sensor
• Range: 0 to 2,000 ppm
• Initial value
b
: 0 to 20 ppm
b.
Sensor exposed to air; for example, during the initial verification or the installation.
HourlyTrend
+xx / 24 h
• Variation in ppm of the gas level for a period in hours
• Period adjustable from 1 to 99 hours
• Initial value
a
: 0 ppm/24 hours (default value of the period)
DailyTrend
+xx / 30 d
• Variation in ppm of the gas level for a period in days
• Period adjustable from 1 to 99 days
• Initial value
a
: 0 ppm/30 days (default value of the period)
(
Date
)
a
(
Time
)
a.
The H201Ti’s clock is set at the factory. Local time may differ from the displayed time; to set it,
see Section 6.2.3 on page 6-8.
Date (year-month-day format)
Time (24-hour format)
Содержание Hydran 201i
Страница 1: ...GE Energy Services Digital Energy Hydran Instruction Manual 201i System ...
Страница 6: ...Instruction Manual vi Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 8: ...Instruction Manual viii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 22: ...Instruction Manual xxii Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 36: ...Chapter 1 Introduction 1 10 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 72: ...Chapter 3 Hydran 201Ci Controllers 3 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 114: ...Chapter 5 Before Installation 5 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 143: ...Installation Part 17997 Rev 6 March 2011 6 29 Figure 6 9 Installing the CPU Module ...
Страница 176: ...Chapter 8 Hydran 201Ci 1 Installation 8 16 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 190: ...Chapter 9 Communications and Networking 9 14 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 216: ...Chapter 11 Operating the Hydran 201i System 11 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 236: ...Chapter 12 Troubleshooting 12 20 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 244: ...Chapter 13 Maintenance 13 8 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 254: ...Appendix B Mechanical Drawings B 2 Rev 6 March 2011 Part 17997 Figure B 3 1 5 In Adaptor ...
Страница 255: ...Hydran 201Ti Intelligent Transmitter Part 17997 Rev 6 March 2011 B 3 Figure B 4 1 In Adaptor ...
Страница 262: ...Appendix C Functional Block Diagrams C 4 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 282: ...Appendix F Default Values of the Parameters F 4 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 284: ...Appendix G Unit Conversions G 2 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 298: ...Appendix I Glossary I 12 Rev 6 March 2011 Part 17997 PAGE LEFT INTENTIONALLY BLANK ...
Страница 299: ......