Appendix B — Humilab chilled mirror hygrometer details
Page
55
Operate mode at the completion of the balance cycle. Alternatively,
press
E
to return to the main menu.
Set RS-232
units
From the main menu press
3
to customize your RS-232C data
output. The units selected for output and the handshake status will
be echoed followed by a sub-menu:
SELECTED UNITS = 1,2
HANDSHAKE = DISABLED
1) DATA 2) HANDSHAKE (TOGGLE) 3) EXIT
Press
1
to set the RS-232C data display. The following sub-menu
will appear:
1) SELECT UNITS 2) USER PROMPT E) EXIT
Press
1
to select the parameters to be displayed. One of the
following sub-menus will appear:
1) DPC 2) DPF 3) RH 4)TMPC
5) TMPF 6) NONE E) EXIT
Press numbers individually to add the associated parameter to the
RS-232C data output. Press
3
(or
6
or
8
, depending on Model) for no
data output or to clear before adding parameters. Press
E
when you
are done.
At the Data sub-menu, press
2
to output data at the prompt of the
user. The following will appear:
STATUS = DISABLED
ALLOW USER TO PROMPT FOR DATA
1) ENABLE 2) DISABLE E) EXIT
Press
1
to Enable User Prompt For Data. Now press
E
repeatedly
until you are back to the main menu. Press
E
to enter Operate
mode. There will not be any data output. Press the following key
sequences followed by
RETURN
to see the parameter associated
with the key sequence:
D1 (DPC)
D2 (RH)
D3 (TMPC)
Please note that the “D” must be uppercase
Содержание HumiLab
Страница 43: ...Chapter 5 Maintenance and Troubleshooting Page 39 Figure 13 Mirror Check and Mirror Sensor Balancing Procedure ...
Страница 75: ...Appendix D Humidity Equations and Conversion Charts Page 71 Figure 16 Graphical Conversion Chart ...
Страница 76: ...Page 72 Appendix D Humidity Equations and Conversion Figure 17 Psychrometric Chart ...
Страница 80: ...Page 76 Appendix E Warranty information and return ...