Appendix A — Specifications
Page
43
Appendix A — Specifications
% RH range
10 to 90% RH at 77°F (25°C)
Temperature range
68 to 86°F (20 to 30 °C) May be used with Tempera-
ture-Controlled Water Bath/Circulator. Humidity limited by
surface temperature of internal walls and cover (dew point
in chamber must be less than interior surface temperature
to avoid condensation).
Response time
10 minutes for 63% step change. 30–40 minutes to full
stability
Accuracy
±1% RH from 10% to 70% RH
±1.5% RH from 70 to 90% RH
±0.3°F (0.15°C) dew point at 77°F (25°C)
±0.3°F (0.15°C) dry bulb at 77°F (25°C)
Analog outputs
Two (humidity and temperature), each user-scalable,
4–20mA/0–5VDC
Power
115 or 230 VAC ±10%, 50/60Hz single phase
Digital interface
Two RS-232 ports for reference and generator interface
Prostep software
Windows 95/98 or above required. Supplied on CD ROM
with Operator’s Manual (PDF) and HCON Humidity Conver-
sion Software.
Approvals
CE approved. Certifications supplied: Certificate of NIST
traceability and functional test data
Display
Three LEDs, 0.5" (1.3cm) high, 7-segment
Shows Set Point (%RH), Actual % RH, and Temperature
0.1% RH/°C/°F resolution
Workspace dimensions
11" by 9" by 6.5" (28 by 23 by 16.5 cm)
Approx. 644 in
3
(10.6 liters)
Overall dimensions
23"L by 19"W by 13"H (58 by 48 by 33 cm)
Содержание HumiLab
Страница 43: ...Chapter 5 Maintenance and Troubleshooting Page 39 Figure 13 Mirror Check and Mirror Sensor Balancing Procedure ...
Страница 75: ...Appendix D Humidity Equations and Conversion Charts Page 71 Figure 16 Graphical Conversion Chart ...
Страница 76: ...Page 72 Appendix D Humidity Equations and Conversion Figure 17 Psychrometric Chart ...
Страница 80: ...Page 76 Appendix E Warranty information and return ...