Service Manual

GE Healthcare

Portascan Service and Assembly Manual
MAN-004-049 
Issue 1

Содержание Healthcare Portascan

Страница 1: ...Portascan Portable Bladder Scanner Service Manual GE Healthcare Portascan Service and Assembly Manual MAN 004 049 Issue 1 ...

Страница 2: ...tic Output information 2 Chapter 4 Connector overview Connections 3 Main board 3 ETX Board 3 Chapter 5 Planned Maintenance Handling and Care 4 Handling and Care probes 4 Agents and procedures that may damage the probes 4 Chapter 6 Unit Disassembly Images 5 Chapter 7 Faults and Solutions Screen Related 7 Probe Related 7 Keypad Related 7 Software Related 7 Chapter 8 Parts list 8 Chapter 9 ESD What i...

Страница 3: ...ws arrowed 5 Figure 10 Note location of long screw 5 Figure 11 Probe connector lock nut 5 Figure 12 Use special Lemo wrench to loosen probe connector nut 5 Figure 13 Cut or remove security sticker then hinge top away to separate case 5 Figure 14 The bottom case and handle grip can be placed to one side 5 Figure 15 Two screws hold the main assembly in place 5 Figure 16 Lift main board peeling away ...

Страница 4: ... boards 6 Figure 22 to reveal ETX Heatsink screws Retain Heatsink from other side 6 Figure 23 The compact flash can now be removed reprogrammed replaced 6 Figure 24 RAM can also be accessed if necessary Observe E S D precautions 6 Figure 25 Thermal Printer PCB and location 6 Figure 26 Paper roll installed 6 Dual Frequency prove connector 6 Rear panel connections 6 Bottom case label 6 Keypad membra...

Страница 5: ...st be carried out by authorized persons The electrical installation of the relevant room must comply with the IEC requirements The product must be used in accordance with the instructions for use Precautions Cleaning the probe is done by first removing the ultrasound coupling gel with a soft tissue and then gently wipe the probe dry using a new tissue or dry cloth When more cleaning is required on...

Страница 6: ...imensions WxDxH 16 x13 x 13 cm Weight 2540g without battery and probe3220g with battery and probe Enclosure Leakage and Earth Leakage Current Within specification for Class II type BF according to EN 60601 1 Environmental Operating Conditions Temperature 8 40ºC Humidity up to 90 20ºC General Storage Conditions Temperature 0 50ºC Standards EN 60601 1 Class II Type BF handheld equipment Acoustic Out...

Страница 7: ...Peripherals Repairable Parts Serviceable by Module Only Figure 1 Portascan Block Diagram Figure 2 Mainboard Top View Figure 3 ETX Board Block Diagram Chapter 4 Connector overview Connections Main Board See Main Board Schematic for further details of connector Pin assignments EXT Board CPU Geode GX1 1028aseT Ethernet Controller Companion Chip CS5530A DRAM BIOS flash memory USB0 USB1 Sound Codec Har...

Страница 8: ...afety of the probe could be compromised Disconnect the probe from the scanner before cleaning and disinfection Make sure to switch the system off before disconnecting or connecting the probe To prevent hazards refer to your local requirements for adequate electrical installation in case of class 1 type BF Agents and procedures that may damage the probes Some agents and procedures damage probes Use...

Страница 9: ... 6 Unit Disassembly Figure 12 Use special Lemo wrench to loosen probe connector nut Figure 13 Cut or remove security sticker then hinge top away to separate case Figure 14 The bottom case and handle grip can be placed to one side Figure 15 Two screws hold the main assembly in place Figure 16 Lift main board peeling away earth strap disconnect plugs skts Figure 17 Note location of sounder fan and d...

Страница 10: ... side Heatsink screws Figure 23 The compact flash can now be removed reprogrammed replaced Figure 24 24 RAM can also be accessed if necessary Observe E S D precautions Figure 25 The compact flash can now be removed reprogrammed replaced Figure 26 24 RAM can also be accessed if necessary Observe E S D precautions LCD mounting brackets screws Dual frequency probe connector Rear panel connections Bot...

Страница 11: ... of the keypad ribbon cable Reconnect or replace keypad Software Related On the rare occasion that Windows 98SE crashes the display will freeze and keypad will be inactive In this situation the system will automatically shut down after 15 seconds DO NOT remove the battery pack until this has happened Removing the battery pack before complete shutdown could cause the following problems Unit boots i...

Страница 12: ...0168B Interface PCB PA00143A Inverter Cable C00049 Screening Cans PSA00062 Keyboard Membrane PSA00182 LCD Backlight Inverter C00048 LCD Bracket Bottom PSA00015 LCD Bracket Top PSA00014 LCD Data Cable PSA00017 LCD Screen C00047 Lid PSA00006A Magnet Contact PSA00061 Magnet Support Plates PSA00060 Handle Soft Grip PSA00011 Magnet Print Lid C00216 Main Case Lower C00037A Main Case Upper C00036A Main P...

Страница 13: ...op Wear ESD safe clothing Always check wrist strap and other ESD equipment before use Be grounded BE AWARE ESD safe workshop To be sure that there is no potential difference between body boards and work surface the work surface should be made of electrostatic dissipative rubber The work surface should be connected to a central ground point The engineer handling the boards should also be connected ...

Страница 14: ...Section 1 Circuit Diagrams 10 Mediwatch Service Department Service Documentation ...

Страница 15: ...Section 1 Circuit Diagrams 11 Mediwatch Service Department Service Documentation ...

Страница 16: ...o select the Set Date option Check that the correct date is displayed If the incorrect date is displayed then set the correct date by using the UP DOWN and TAB menu buttons and lock the setting by pressing the SELECT button Press the menu button Using the UP DOWN menu buttons select the Date Time option and then press the SELECT button Press the DOWN button and then the SELECT button to select the...

Страница 17: ...f the result On the printed results record in the field Patient Name the serial number of the unit that is being inspected Open the printer door and raise the print head and remove the paper Feed the paper back into the printer until the end appears from the front paper slot and re engage the print head Press the print button The last recorded image should be printed Leave the unit sitting for 5 m...

Страница 18: ...ns and features shown herein or discontinue the product described at any time without notice or obligation Visit us online at www gehealthcare com 2005 General Electric Company All Rights Reserved For more than 100 years scientists and industry leaders have relied on General Electric for technology services and productivity solutions So no matter what challenges your healthcare system faces you ca...