8.17.1 Mounting/Replacing a Secondary "Patient" Monitor
Note
A Secondary Monitor is NOT intended for diagnostic use. It is an additional device used to allow the patient
to watch the proceedings.
8.17.1.1 Preparing the Isolation Transformer
Figure 8-31 Isolation Transformer kit for Secondary Monitor
Caution
For changing the input/output voltage or fuses the isolation transformer should not be connected. The wrong
fuses and position of the input/output voltage selector may cause major damage on connected peripherals.
1.
Before using the Isolation Transformer you must check the input/output voltage settings to meet the
ratings of the line power available in your location or country.
2.
For changing the input/output voltage, remove the label by means of a small socket wrench.
3.
Switch both voltage selectors (input as well as output voltage) to 115V or 230V. Consider that changing
the input/output voltage also requires to change the fuses!
Figure 8-32 change input/output voltage
4.
For changing the fuses, open the fuse protection at the power inlet block with a small screwdriver,
remove the fuse holder and exchange fuses.
Figure 8-33 exchange
fuses
115V (100-130V) -> 4AT
230V (220-240V) -> 2AT
5.
Assure that the connected loads can be operated with the chosen voltage.
Replacement Procedures
Voluson E-Series Service Manual
KTD106657 Revision 2
8-27
Содержание H48681XB
Страница 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Страница 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Страница 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Страница 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Страница 36: ...This page was intentionally left blank Introduction 1 28 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 172: ...This page was intentionally left blank Functional Checks 4 46 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 226: ...This page was intentionally left blank Service Adjustments 6 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 288: ...This page was intentionally left blank Replacement Procedures 8 30 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 364: ...This page was intentionally left blank Care and Maintenance 10 24 Voluson E Series Service Manual KTD106657 Revision 2 ...
Страница 365: ......
Страница 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...