
GE H
EALTHCARE
RAFT
V
OLUSON
E8 / V
OLUSON
E6
D
IRECTION
KTD102576, R
EVISION
7
DRAFT (A
UGUST
23, 2012)
S
ERVICE
M
ANUAL
3-26
Section 3-5 - Connection of Auxiliary Devices
Connecting the 19” LCD Secondary “Patient” Monitor
Figure 3-20 Connection Scheme - 19” Secondary LCD Monitor
text
Console back
DVD Writer
I
O
S - Video
Out
In
Out
DVI - Out
A
BC
DE
Lift
is only used for the lifter
(
230V
- can not be changed)
19” LCD
Secondary Monitor
appropriate
VGA cable
connected to the
rear side of the
2nd Monitor
Transformer
A Secondary “Patient” Monitor
MUST NEVER
be connected to the Voluson E8 / Voluson E6 ultrasound
systems mains supply directly!
Always connect it to an appropriate
Isolation Transformer
(see:
Chapter 9 - Optional Equipment, on page 9-33
CAUTION
!! CAUTION: