GE H
EALTHCARE
DRAFT
V
OLUSON
E8 / V
OLUSON
E6
D
IRECTION
KTD102576, R
EVISION
7
DRAFT (A
UGUST
23, 2012)
S
ERVICE
M
ANUAL
Chapter 3 - Setup Instructions
3-23
3-5-7
Connecting the VGA Image (Video) Resizer
NOTE:
After physical connection of the VGA Image Resizer to the Voluson system, follow the procedure
described in
Section 3-5-7-1 "Adjustment of the VGA Image Resizer Settings" on page 3-24
.
NOTICE
!! NOTICE:
For more detailed description, how to replace/mount the VGA Image Resizer, refer to:
Chapter 8 - Replacing optional Peripherals / How to mount Peripherals at a later date, on page 8-25
NOTICE
!! NOTICE:
The VGA Image (Video) Resizer is required whenever the used Secondary “Patient” Monitor has a
different screen resolution than the Voluson E8 / Voluson E6 system!
Figure 3-18 Connection Scheme - VGA Image Resizer
text
Console back
DVD Writer
I
O
S - Video
Out
In
Out
DVI - Out
A
BC
DE
Power Supply
VGA Image
to Secondary
“Patient” Monitor
Lift
is only used for the lifter
(
230V
- can not be changed)
CAUTION
!! CAUTION:
A Secondary “Patient” Monitor
MUST NEVER
be connected to
the Voluson E6 / Voluson E8 systems mains supply directly!
Always connect it to an
appropriate Isolation Transformer
(see:
Chapter 9 - Optional Equipment, on page 9-33
Resizer box
Connect the power supply of the VGA Image Resizer to the
power cable plug behind the door on the right side cover.