8-24
Section 8-16 - Replacement of the Monitor Transportation Lock (BT09 only)
Section 8-16
Replacement of the Monitor Transportation Lock (BT09 only)
NOTE:
The Monitor Transportation Lock is only used if the “outdated” monitor arm (KTZ154817) is mounted on
the Voluson system.
8-16-1
Manpower
One person, 2 minutes
8-16-2
Tools
Phillips screwdriver PH1 and PH2
8-16-3
Monitor Transportation Lock - Removal Procedure
1.) Unscrew 1 screw, then remove the Monitor Transportation Lock.
8-16-4
Monitor Transportation Lock - Installation Procedure
1.) Attach the new Monitor Transportation Lock and fasten it with 1 screw.
Figure 8-26 remove 1 screw
1 screw
Monitor
Transportation Lock