8-60
Section 8-4 - System Modules- Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-4-11
T-CFE (cFront End) Module Replacement Procedure
8-4-11-1
Tools
Appropriate Phillips screwdriver.
8-4-11-2
Time Required
15 minutes
8-4-11-3
Preparation
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-11-4
T-CFE Module Removal Procedure
1) Remove all accessories.
2) Remove the following covers: RS Probe Cover, Lower Front eTower, Left side eTower, and Right
Side eTower.
3.) Open and remove the full Front End door assembly and place it on a stable surface:
4.) Remove the T-FEPS module.
CAUTION
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
FRU Part #
Refer to
.
•
Accessories - Replacement Procedures
•
RS Probe Cover Removal Procedure
•
Lower Front eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
•
Содержание H45051VN