Chapter 8 - Replacement Procedures
8-21
8-2-13
Rear Handle Replacement Procedure
8-2-13-1
Tools
Use the appropriate Phillips-type screw drivers and an 8mm Allen wrench, to replace the rear handle.
8-2-13-2
Preparation
Shut down the Vivid™ 4 ultrasound unit, as described in
.
8-2-13-3
Rear Handle Removal Procedure
1) Remove the control console lower cover, as described in the
2) Remove the control console upper cover rear, as described in the
3.) Remove the eight Allen 8mm securing screws (four on each side) from the rear of the upper
console.
4) Remove the Rear Handle.
8-2-13-4
Rear Handle Installation Procedure
1) Return the Rear Handle to its original position on the rear of the machine.
2) Fasten each side of the Rear Handle to the rear of the upper console with its 8 mm securing nuts
(eight in total).
3) Return the control console upper cover rear to its original position, as described in the
Console Upper Cover (Rear) Installation Procedure
4) Return the control console lower cover to its original position, as described in the
Figure 8-20 Rear Handle