4. Divide the lowest reading by the highest reading. If the LED light is functioning properly, the result
should be greater than 0.4. Refer to symptom S12 in Chapter 10 for troubleshooting.
8.4.4
Electrical Safety Tests
It is recommended that the electrical safety tests be performed as follows:
• Before the first use of Giraffe Blue Spot PT Lite
• Each time the system is serviced
• Every twelve months thereafter
Use an approved electrical safety analyzer. Follow the operating instructions supplied by the vendor of the
electrical safety analyzer. For reliable leakage current checks, perform the ground resistance check first.
8.4.4.1
Ground Resistance Check
The ground resistance check varies depending on the power source:
• For a system plugged into a wall outlet, measure the resistance between power cord ground terminal
and the potential equalization stud on the system. The ground resistance must be less than 0.1 ohms.
• For a system plugged into an accessory outlet on a microenvironment, measure the resistance
between the power cord ground terminal of the microenvironment and the potential equalization stud
on the Giraffe Blue Spot PT Lite. The ground resistance must be less than 0.2 ohms.
8.4.4.2
Leakage Current Checks
In normal conditions and in all possible operating modes:
• For a system plugged into a wall outlet, measure the leakage current and confirm that it is less than
130 microamperes.
• For a system plugged into an accessory outlet on a microenvironment, measure the leakage current
and confirm that it does not increase leakage current of the microenvironment by more than 130
microamperes.
If required by local ordinances, in single fault condition and in all possible operating modes:
- For a system plugged into a wall outlet, measure the leakage current and confirm that it is less than 500
microamperes.
- For a system plugged into an accessory outlet on a microenvironment, measure the leakage current
and confirm that it does not increase leakage current of the microenvironment by more than 500
microamperes.
54
Chapter 8: Service Maintenance and Checkout
Содержание Giraffe Blue Spot PT Lite
Страница 1: ...Giraffe Blue Spot PT Lite Operation Maintenance and Service Manual GE Healthcare ...
Страница 7: ... 7 Part I Operation and Maintenance ...
Страница 8: ...8 Part I Operation and Maintenance This page is intentionally left blank ...
Страница 16: ...This page is intentionally left blank 16 Chapter 1 Product Overview ...
Страница 22: ...This page is intentionally left blank 22 Chapter 2 Product Setup and Operation ...
Страница 26: ...This page intentionally left blank 26 Chapter 3 Operator s Maintenance ...
Страница 32: ...This page intentionally left blank 32 ...
Страница 33: ... 33 Part II Service ...
Страница 34: ...34 Part II Service This page is intentionally left blank ...
Страница 44: ...44 Important Service Safety Information This page is intentionally left blank ...
Страница 50: ...This page is intentionally left blank 50 Chapter 7 Installation ...
Страница 55: ...Chapter 9 Calibration No calibration is required for the Giraffe Blue Spot PT Lite 55 ...
Страница 56: ...This page is intentionally left blank 56 Chapter 9 Calibration ...
Страница 72: ...This page is intentionally left blank 72 Chapter 11 Replacement Procedures ...
Страница 88: ...12 5 Wiring Diagram 88 Chapter 12 Service Parts ...
Страница 94: ...This page intentionally left blank 94 ...
Страница 95: ......