GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
8 - 288
Section 8-11 - Back End Processor (BEP) parts replacement
8-11-13
DVI Flex Main Yggdrasil replacement
8-11-13-1
Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the VIVID E9.
2.) Disconnect the Mains Power Cable from the wall outlet.
3.) Disconnect all probes.
4.) Remove the Left Side Cover.
5.) Remove the
BEP Cover
.
a.) Loosen the BEP Cover’s fixing screw(s).
-
On BEP6 there are two finger screws on the top of the BEP Cover.
-
On newer BEP5 there are three Phillips screws, one on the top and one on each side of
the BEP Cover.
-
On older BEP5 there is one finger screw at the top of the BEP Cover.
b.) Tilt the top of the
BEP Cover
away from the
BEP
.
c.) Set the
BEP Cover
aside.
6.)
Related information:
•
4-2-2 "Power shut down" on page 4-8.
•
8-5-2 "Side Covers replacement" on page 8-38.
8-11-13-2
DVI Flex Main Yggdrasil removal
1.) Disconnect one end of the
DVI Flex Main Yggdrasil
from the connector on the Graphics Adapter.
2.) Disconnect the other end of the
DVI Flex Main Yggdrasil
from the DVI-IN connector on the DVR.
8-11-13-3
DVI Flex Main Yggdrasil installation
1.) Install one end of the
DVI Flex Main Yggdrasil
to the DVI-IN connector on the DVR.
2.) Install the other end of the
DVI Flex Main Yggdrasil
to the connector on the Graphics Adapter.
3.) Install the
BEP Cover
.
4.) Install the
Left Side Cover
.
Содержание GA000100
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5 ...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO ...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii ...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12 ...
Страница 801: ......