GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
4-12
Section 4-2 - General Procedures
4-2-5
Using Removable Media
4-2-5-1
Introduction
Vivid q N is equipped with a Magneto Optical Disk Drive (MOD Drive) and a DVD/CD-RW Drive (as an
option from system version BT12).
The MOD Drive can be connected to either of the USB ports (
upper
or
lower
) on the rear connectors
panel. This connection must be via a USB-to-SCSI adaptor; one side of the adaptor being connected to
the Vivid q N USB port, and the other side to the rear of the MOD drive.
When connecting the MOD drive directly to the system, it is necessary to use an additional power
source.
As a safety precaution, this must be done via an isolation transformer.
The MOD Drive may be connected once the system is powered ON, or after shutdown. All software
drivers for the MOD Drive are pre-configured and installed designated to be used with the specific MOD
drive supplied by GE.
WARNING
WARNING
Do not attempt to use a different type of mass storage device other than the
MOD Drive supplied with the system. The ultrasound system is an extremely sensitive
and complex medical system. Any unauthorized peripherals may cause system failure
or damage!
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...