5.3.4
Isoelectric focusing from the control panel
Precautions
WARNING
Ettan IPGphor 3 is a high voltage instrument that can cause fatal
electrical shock if the safety features are disabled. The safety lid
must be securely closed before starting a protocol.
Protocol steps
Once the protocol begins, each step is carried out automatically, and instrument status
for each step is indicated on the screen. Platform temperature and time remaining can
be changed while the protocol is in progress, and each such possibility is discussed in
turn below.
Select protocol number
Action
Step
Place the cursor in the protocol number field using the Right and Left arrow
keys and select the protocol number using the Up and Down arrow keys.
1
P r o t # 1 M Y P R O T O C O L
R e h y d r a t e 1 0 : 0 0 H r @ 2 0
˚
C
I E F @ 2 0
˚
C 2 0 0
µ
A / S t r i p
1 2 S t r i p s 3 S t e p s
Use the Right and Left arrow keys to position the cursor to set the number
of strips to run, and the up and Down arrow keys to select the number(1–12)
2
P r o t # 1 M Y P R O T O C O L
R e h y d r a t e 1 0 : 0 0 H r @ 2 0
˚
C
I E F @ 2 0
˚
C 2 0 0
µ
A / S t r i p
1 2 S t r i p s 3 S t e p s