5.3.2
Protocol parameters
Programmable parameters
Ettan IPGphor 3 can store up to ten protocols. Programmable protocol parameters include:
•
Duration and temperature of the rehydration phase.
•
Duration, temperature, and power parameters for up to nine IEF steps for each pro-
tocol.
•
Power parameters including maximum current, voltage, and voltage change pattern.
Please refer to the 2-D Electrophoresis, principles and methods guide for sample handling
recommendations and protocol guidelines. The permitted range of values for each pro-
grammable parameter is listed below.
Permitted range of values for each programmable parameter
Recommended values
Operational Range
Parameter
1 to 9
Steps
10:00 to14:00 hrs
Rehydration: 0 to 99:59 hrs
Step duration
2:00 to 10:00 hrs
Focusing: 0 to 99:59 hrs
100 to 10000 V
0 to 10000 V
Volts
50–75 μA per strip
1500 μA total for 12 strips max.
Max. current
20
Rehydration: 15°C to 30°C
Platform temperature
20
Focusing: 15°C to30°C
Ettan IPGphor 3 Operating Instructions 28-9635-51 AC
73
5 Operation
5.3 Perform a run from the control panel
5.3.2 Protocol parameters