User Manual
Druck DPI612
Page 50 of 105
[EN] English – 109M4017 Revision -
2.3.5
Applying Hydraulic Pressure (1000 bar)
1
2
3
4
Step
Procedure
1
To seal the system
2
Close the refill valve and then wind the volume adjuster fully
clockwise and counter clockwise until the pressure starts to
increase.
Continue the clockwise/counter clockwise sequence until you
get the necessary pressure OR for full control, go to step 3.
The counter clockwise operation refills the pressure
mechanism but there is no change in pressure to the device
under test or the PM 620 module (if applicable).
Note:
At higher pressures, it is easier to turn the wheel if you
fold in the handle; see Section 1.4.5.
3
For full control, open the refill valve (1 turn). You can now
increase (+) or decrease (-) the pressure with the volume
adjuster.
4
If you increase pressure and get to the limit of travel, close
the refill valve again and wind the volume adjuster fully
counter clockwise.
5
Continue to do steps 2 to 4 until you get the necessary
pressure