User Manual
Druck DPI620 Genii
Table 8-1
Function
Calibration equipment
(ppm = parts per million)
Current
(CH1 or CH2)
Current (mA) calibrator.
Accuracy - Current measure/source,
refer to Table 8-2 or Table 8-3
Voltage
(CH1 or CH2)
Volts calibrator.
Accuracy - Voltage measure/source,
Refer to Table 8-5 or Table 8-7
Millivolts
(CH1 or CH2)
OR
TC mV (CH1)
mV calibrator.
Accuracy - Millivolts measure/source,
refer to Table 8-4 or Table 8-6
Accuracy - TC mV refer to Table 8-14
Frequency
(CH1)
measure
source
Signal generator
Total error: 0.3
ppm or better
Frequency meter
Total error: 0.3
ppm or better
Resolution: 8
digits (minimum)
Resistance
(CH1)
measure
source
Standard resistor
100R, 200R, 300R,
400R, 1k, 2k, 4k
Total uncertainty:
20 ppm
An ohmmeter or an
RTD measurement
system with the
specified excitation
currents, refer to
Cold Junction
(CH1)
Calibrated K type thermocouple
Accuracy: 50 mK for -5 to 28°C (23 to
82.4°F)
06 June 2013
K0541 issue 1
Page 140 of 175