Preferences
6.13
Select the Preferences icon to setup the HART ® device
search method.
The application allows searching by:
•
Poll Address - when each transmitter has a unique
address.
•
Short tag - if the transmitter supports 8 character tags.
•
Long tag - if the transmitter supports 32 character
tags.
Transmitters with a non-zero poll address are in multi-drop
mode and default to a fixed loop current of 4 mA.
By default, the Druck DPI 620 Genii polls address 0 (zero)
only. Change the Poll Address by selecting the appropriate
search radio button or entering the tag name in the search
field.
06 June 2013
K0541 issue 1
Page 107 of 175